In 1999, the Institute of Primary Metals and the Institute of Corrosion and Protection of Primary Metals were integrated to establish a new "Institute of Metal Research of the Chinese Academy of Sciences" and entered the "Northeast High Performance Materials Research and Development Base" of the Knowledge Innovation Project of the Chinese Academy of Sciences.

Slow strain rate (SSRT) stress corrosion (SCC) testing machine is used for corrosion tensile test of materials in corrosive environment.

Technical parameters of slow stress corrosion testing machine:

1. Maximum test force: 50kN, 100kN

2. Test force range: 1%~100%F.S.

3. Relative error of test force indication: ±0.5%

4. The maximum moving range of the loading head: 80mm

5. Moving speed of loading head: 1mm/s~1x10-7mm/s

6. Measuring accuracy of loading head displacement: ±0.5%

7. Loading head displacement resolution: 0.0125µm

8. Deformation range of sample: 0~10mm

9. Deformation measurement range: 0~30mm

10. Deformation measurement resolution: 1µm

11. Deformation measurement accuracy: ±0.5%

12. Test temperature: normal temperature~550℃

13. Temperature control accuracy: ≤±1℃

Institute of Metal Research, Chinese Academy of Sciences (1)